AGILENT Compatible Instruments

AGILENT INSTRUMENTS
OEM Applications
Application
Requirements
Solution
Agilent 1100 Series LC/MSD
Nitrogen up to 15L/min, 80 - 100 psi
iGen N2-35
(At least 3L/min is required at all times to prevent air entering the instrument)
Agilent 1200 Series HPLC with ELS Detector
Nitrogen up to 4L/min, 50 psi (for G4218A ELSD)
iGen N2-35
Agilent 385 ELSD
Nitrogen 0.9 - 3.25 L/min @ 60 - 100 psi
iGen N2-35
Agilent 6100 Series
Nitrogen up to 15L/min, 80 – 100 psi
iGen N2-35
(At least 3L/min is required at all times to prevent air entering the instrument)
Agilent 6200 Series
Nitrogen up to 18L/min, 80 – 100 psi
Agilent 6300 Series Ion Trap LCMS
Agilent 6400 Series 6410 / 6420/ 6430
Nitrogen up to 18L/min, 80 – 100 psi
iGen N2-35
(At least 3L/min is required at all times to prevent air entering the instrument)
Agilent 6460 /6475A
Nitrogen up to 30L/min, 80 – 100 psi
(At least 3L/min is required at all times to prevent air entering the instrument)
Agilent 6490 LCMS QQQ iFunnel
Nitrogen up to 50L/min, 80 – 100 psi
iGen N2-50
(At least 9L/min is required at all times to prevent air entering the instrument)
Agilent 6495D LC/TQ
N2 up to 50LPM; with 9LPM minimum
iGen N2-50
Agilent CHIP CUBE
Up to 30 L/min of nitrogen standard + Requires 4 L/min of air to be mixed in with the nitrogen supply to achieve 10% oxygen content for Chip Cube operation.
TBD
Chip Cube is available on all 6000 series LC/MS models.
CHIP CUBE on iFunnel models 6490 and 6550
Up to 60 L/min of nitrogen standard. + Requires 10 L/min of air to be mixed in with the nitrogen supply to achieve 10% oxygen content for Chip Cube operation
TBD
Agilent 6500 Series QTOF LC/MS
Nitrogen up to 20 L/min for 6510, 6520, 6538 Nitrogen up to 30 L/min for 6530 and 6540
(At least 3L/min is required at all times to prevent air entering the instrument)
iGen N2-35
6550 iFunnel Q-TOF LC/MS
Nitrogen up to 60 L/min for 6550
iGen N2-50
(At least 12L/min is required at all times to prevent air entering the instrument)
Agilent models available with Jet Stream technology:
Nitrogen 30 L/min @ 100 psi
iGen N2-35
Nitrogen for Plasma Operation:
22.5 L/min at 72.5 psi (99.5%+)
Agilent 4100 MP-AES
Nitrogen for Optics purge:
To be advised.
10 L/min at 65 psi (99.95%+)
(Agilent offer own-brand 4107 PSA Nitrogen generator at the moment)
Microwave Plasma-Atomic Emission Spectrometer
Clean, Dry, Oil-free Compressed Air for:
Preoptics protection: 25 L/min at 72.5 psi
iGen ZA-36000
Optional ECGM: 1.5 L/min at 72.5 psi
iGen ZA-1500
Agilent 70 series ICP-OES
High Purity Nitrogen
NG Range
0.45 L/min in standby mode
Up to 2.1 L/min in operational mode
Agilent MP-AES
10L/min of Nitrogen 99.5% Purity 36.5L.min of Dry Air
iGen QDS
Agilent G6465A/B Ultivo LC/TQ
18 lpm (ESI)
iGen N2-35
30 lpm (AJS)
iGen N2-50