Thermo SCIENTIFIC Compatible Instruments
THERMO SCIENTIFIC INSTRUMENTS
OEM Applications
Application
Requirements
Solution
Thermo** LTQ Range
Nitrogen up to 21 L/min at 100 psi
iGen N2-35
Thermo Quest LCQ DECA XP +
Nitrogen up to 30 L/min at 100 psi
THERMO LTQ VELOS ETD
Nitrogen up to 15 L/min 100 psi
VELOS PRO ETD
LTQ XL ETD
ETD option: 1cc/min 99.999% N2 @ 40 psi
iGen N2-35
THERMO LTQ ORBITRAP VELOS With ETD Module (Electron Transfer Dissociation)
Nitrogen for the API source and C-Trap: 99.5% pure, flow rate 15 L /min at 100 ± 20 psi
Nitrogen supply for ETD option: UHP 99.999% w/less than 1 ppm H2O & O2. Pressure at:40 ±5 psi
iGen N2-35
ETD module consumes less than 1 ATM cc/m of reagent carrier gas, so a standard large bottle (245 ft3) of gas should last approximately two years.
Thermo LTQ Series
Nitrogen up to 21 L/min 100 psi
LTQ XL / LTQ Orbitrap Velos
Thermo Finnigan
Nitrogen up to 21 L/min at 100 psi
iGen N2-35
TSQ Quantum
TSQ Quantum Ultra
TSQ Quantum Discovery
Thermo Exactive
Nitrogen up to 30 L/min at 100 psi +- 10 psi N
iGen N2-35
Thermo Q- Exactive
Nitrogen up to 30 L/min at 116 psi
iGen N2-35
Thermo Q- Exactive Plus
Thermo LXQ Linear
Nitrogen up to 15 L/min, up to 100psi
iGen N2-35
Thermo LCMS with HESI or HESI-II probe attachment.
Typical settings, as recommended by Thermo do not exceed 25 L/min, however, at max gas settings with HESI probe, nitrogen flow requirement was measured up to 40L/min, 100 psi.
iGen N2-35
Please note; Thermo states that these settings are unlikely to be used for real analysis as they are not optimum and would degrade the signal for any given sample. Verify with customer.
iGen N2-50
Thermo/ Dionex*** MSQ+
Nitrogen up to 15L/min, 100 psi
iGen N2-35
Thermo 51i Total HydroCarbon Analyzer with FID
Makeup Air Flow Rate:
iGen ZA-1500
150 - 300 cc/min Hydrocarbon free air
Fuel Flow Rate:
10 to 35 cc/min Hydrogen
N/A
Thermo 55C Hydrocarbon Analyzer
Support Gas Pressures
Zero Air: 20 - 30 PSI
iGen ZA-1500
Hydrogen: 20 - 30 PSI (hydrogen)
N/A
UHP Zero Nitrogen: 25 - 35 PSI
HC Free Air (200 - 300 cc/min)
iGen ZA-1500
N2 carrier (35 cc/min)
H2 Fuel (25 cc/min)
N/A
Note: additional pressure regulator required!
Thermo Trace GC/FID
Carrier gas: He, H2 or N2
N/A
Capillary Column: Typically 2ml/min
Packed Column: 40ml.min
Make up gas: N2 or He
10-60ml.min / Typically 30ml.min
(only used with capillary columns)
Detector gas:
H2 for fuel gas 30-50ml/min
N/A
ZA for flame support 300-600ml/min
iGen ZA-1500
*** Thermo/ Dionex MSQ:
• N118LA model number is used for general sales
• N418LA model number is ONLY used for direct sales to Thermo/ Dionex.
• N418LA is rebadged by Dionex to MSQ18LA