WATERS Compatible Instruments
WATERS INSTRUMENTS
OEM Applications
Application
Requirements
Solution
Waters LCT Premier Range
Nitrogen up to 30L/min, 100psi
Waters ZQ /EMD
Waters Acquity Range
Waters 3100 Mass Detector
iGen N2-35
Waters LCT Premier Range
Waters ZQ /EMD
Waters Acquity Range
Waters 3100 Mass Detector
Waters Synapt Range
API gas: Nitrogen up to 20L/min, 100psi
Synapt G2 HDMS also requires IMS gas: >99.5% N2 @ 14 psi, 10 – 100 ml/min
iGen N2-35
*Common gas supply for API & IMS is ok if higher purity is used.
Waters XEVO TQS
Nitrogen at 23.5 L/min, 100psi
Waters G2 QTOF
Waters QTOF API-US
Desolcation Gas: N2 @ 3 L/min - 12.5 L/min
iGen N2-35
API-LC/MS/MS
Cone Gas: N2 @ 1 L/min - 5 L/min
Nebulizing Gas: 0.3 L/min
Max Total N2 18 L/min at 90-100 psi
Waters QTOF API-US
Desolvation Gas: N2 @ 3 L/min - 12.5 L/min
API-LC/MS/MS
Cone gas: N2 @ 1 L/min – 5 L/min
Nebulizing gas: 0.3 L/min
Max Total N2: 18 L/min at 90 – 100 psi
iGen N2-35
Waters Quattro Micro LC/MS/MS
Nitrogen up to 20L/min, 100psi
Waters Quattro Premier
Nitrogen up to 25L/min, 100psi
Waters Quattro Micro GC/MS/MS
Nitrogen up to 20L/min, 100psi
Waters QTOF Micro
Nitrogen between 10 – 20 L/min, 100psi
iGen N2-35
Waters QDa
Nitrogen up to 20 L/min @ min 98 psi
iGen N2-35